Instrument Function
R&S
®
SMC100A
98Operating Manual 1411.4060.32 ─ 09
Set to default
Enables you to reset the *IDN and *OPT strings in user defined mode, see "Mode"
on page 97 .
The default strings vary depending on the selected emulation mode (Language)
IDN String
Indicates the identification string of the instrument when queried with the common
command *IDN?.
To assign a user defined identification string, select Mode "User defined".
Remote command:
*IDN? on page 235
OPT String
Indicates the option string of the instrument as queried with the common command
*OPT?.
If a "User defined"(see Mode) IDN String is selected, you can create a user defined
option string in addition to the automatically created one.
Remote command:
*OPT? on page 235
5.2.3.14 Protection
This "Protection" dialog provides access to the unlocking of different protection levels .
Access:
► Select "Setup > Protection"
After power on the instrument, all protection levels are locked. To unlock the protec-
tion, the correct password must be entered, see "To unlock or lock a protection level..."
on page 99.
The following functions are protected in the respective levels:
●
Protection Level 1
Protects against accidental changes to certain settings, e.g. clock and date, net-
work settings or instrument names. You can access this protection level with the
password 123456.
●
Protection Level 2
Provides access to the unlocking of protected service functions. It is accessible for
authorized personnel of Rohde & Schwarz service departments.