Instrument Function
R&S
®
SMC100A
164Operating Manual 1411.4060.32 ─ 09
5.3.7.3 RF Level Sweep
To open the "Level Sweep" menu, select "RF > Configure > Level Sweep" or use the
MENU key under "RF".
RF Level Sweep Settings
In the top section, the RF level sweep is activated and the sweep mode is selected.
The buttons are used to reset the level sweep (all sweep modes) or to execute the
level sweep ("Single" mode).
The sweep range, sweep spacing and dwell time are set in the bottom section.
State - Level Sweep
Activates Level Sweep mode.
Note: Activating the RF level sweep mode automatically deactivates other sweeps.
Remote command:
[:SOURce<hw>]:POWer:MODE on page 318
Mode - Level Sweep
Selects the level sweep instrument operating mode and the sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
"Auto"
Sets an automatically repeated sweep cycle.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR AUTO
SOUR:POW:MODE SWE