Instrument Function
R&S
®
SMC100A
188Operating Manual 1411.4060.32 ─ 09
Remote command:
[:SOURce<hw>]:PULM:WIDTh on page 329
Pulse Delay - Pulse Generator
(External trigger only)
Sets the pulse delay. The pulse delay determines the time that elapses after a trigger
event before pulse modulation starts. The pulse delay is not effective for double pulse
generation.
Remote command:
[:SOURce<hw>]:PULM:DELay on page 325
Double Pulse Width - Pulse Generator
(Double Pulse only)
Sets the width of the second pulse.
Remote command:
[:SOURce<hw>]:PULM:DOUBle:WIDTh on page 326
Double Pulse Delay - Pulse Generator
(Double Pulse only)
Sets the delay from the start of the first pulse to the start of the second pulse.
Remote command:
[:SOURce<hw>]:PULM:DOUBle:DELay on page 326
Trigger Mode - Pulse Generator
Selects the trigger mode for pulse modulation.
Note: An external trigger signal is supplied via the PULSE EXT connector.
"Auto"
The pulse generator signal is generated continuously.
"Ext Triggered"
The pulse generator signal is triggered by an external trigger event.
"Ext Gated"
The pulse generator signal is gated by an external gate signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:MODE on page 329
External Trigger Input Slope - Pulse Generator
(External Trigger only)
Sets the polarity of the active slope of an applied trigger signal.
"Positive"
The pulse generator is triggered on the positive slope of the external
trigger signal.
"Negative"
The pulse generator is triggered on the negative slope of the external
trigger signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe on page 329