When used in conjunction with the commands, the parameter <file_name> is speci-
fied as a string parameter with quotation marks. It can contain either the complete path
including the drive, only the path and the file name, or only the file name. The file name
must include the file extension. The same applies for the parameters
<directory_name> and <path>.
Depending on how much information is provided, either the values specified in the
parameter or the values specified with the command MMEM:CDIR (default directory)
are used for the path and the drive settings in the commands.
Before the instrument settings can be stored in a file, they have to be stored in an inter-
mediate memory using common command *SAV <number>. The specified number is
subsequently used in the :MMEMory:STORe:STATe on page 264 command. Also,
subsequently to loading a file with instrument settings with command :MMEMory:
LOAD:STATe on page 263, these settings have to be activated with the common com-
mand *RCL <number>.
7.10.2 Extensions for User Files
The following table lists all available file extensions for user files. The currently availa-
ble files on the instrument depend on the installed options.
Table 7-1: List of the automatically assigned file extensions in the instrument
Function List type Contents File suffix
Instrument State Settings Instrument settings
*.savrcltxt
"User Correction" List User-defined level correction values
*.uco
Export Data *.txt or *.csv
NRP Settings Settings NRP Settings
*.nrp
7.10.3 Examples
In these examples, the current instrument setting is stored in the file
test.savrcltxt in the directory /var/user/..
Storing and Loading Current Settings
1. Store the current setting in an intermediate memory with the number 4. This setting
can be called using command *RCL and the associated number of the memory, for
example *RCL 4.
*SAV 4
2. To store the settings in a file in a specific directory, specify the complete path.
MMEM:STOR:STAT 4,"/var/user/test.savrcltxt"
3. To store the settings in a file in the default drive, set the default drive and specify
only the file name.