263Operating Manual 1411.4060.32 ─ 09
<HeapUsedKb> integer
Used heap memory.
<HeapAvailableKb> integer
Available heap memory.
Usage: Query only
:MMEMory:LOAD:STATe <SavRclStateNumb>, <file_name>
Loads the specified file stored under the specified name in an internal memory.
After the file has been loaded, the instrument setting must be activated using an *RCL
command.
Setting parameters:
<SavRclStateNumb> Determines to the specific <number> to be used with the *RCL
command, e.g. *RCL 4.
<file_name> String parameter to specify the file name with extension
*.savrcltxt.
Example: See "Storing and Loading Current Settings" on page 257.
Usage: Setting only
Manual operation: See " Recall " on page 114
:MMEMory:MDIRectory <Directory>
Creates a subdirectory for mass memory storage in the specified directory. If no direc-
tory is specified, a subdirectory is created in the default directory. This command can
also be used to create a directory tree.
Setting parameters:
<Directory> string
String parameter to specify the new directory.
Example: See "Working with Files and Directories" on page 258.
Usage: Event
Manual operation: See "Create New Directory" on page 117
:MMEMory:MOVE <SourceFile>, <DestinationFile>
Moves an existing file to a new location or, if no path is specified, renames an existing
file.
Setting parameters:
<SourceFile> string
String parameter to specify the name of the file to be moved.