Remote Control Commands
R&S
®
SMC100A
272Operating Manual 1411.4060.32 ─ 09
Example:
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:LENG 16
sets a filter length of 16. The resulting measurement time is 640
ms (2x16x20 ms).
Manual operation: See "Filter Length" on page 154
:SENSe<ch>[:POWer]:FILTer:NSRatio <NSRatio>
The command defines the noise content for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value determines the
proportion of intrinsic noise in the measured result.
Parameters:
<NSRatio> float
Range: 0.001 to 1
Increment: 0.001
*RST: 0.01
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR 0.2
sets a noise content of 0.2.
Manual operation: See "Noise Content" on page 154
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe <MTime>
The command defines the timeout for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value ensures limited
settling times.
Parameters:
<MTime> float
Range: 1 to 999.99
Increment: 0.01
*RST: 4
Default unit: s
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR .2
sets a noise content of 0.2.
SENS1:FILT:NSR:MTIM 5
limits the settling time to 5 seconds
Manual operation: See "Timeout" on page 154