Remote Control Commands
R&S
®
SMC100A
275Operating Manual 1411.4060.32 ─ 09
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
SENS1:POW:OFFS 0.4dB
sets a level offset of 0.4 dB
SENS1:POW:OFFS:STAT ON
a level offset of 0.4 dB is added to the measured value.
Manual operation: See " Level Offset " on page 153
SENSe<ch>[:POWer]:SNUMber?
The command queries the serial number of the sensor.
Return values:
<Snumber> string
Example:
SENS:SNUM?
queries the serial number.
Usage: Query only
Manual operation: See "Current Sensors" on page 87
:SENSe<ch>[:POWer]:SOURce <Source>
The command selects the signal source for the measurement.
Parameters:
<Source> A | B | USER | RF
*RST: A
Example:
SENS:SOUR A
selects the RF signal as measurement source. The RF fre-
quency is used as the measurement frequency of the sensor
and the corresponding correction factor is used. The level set-
ting of the instrument serves as reference level of the measure-
ment.
Manual operation: See " Source " on page 152
SENSe<ch>[:POWer]:STATus[:DEVice]?
The command queries if a sensor is connected to the signal generator.
The sensor is selected by suffix in the keyword SENSe or READ of the command
header. Suffix 1 denotes the sensor connected to the SENSOR connector, suffix 2 the
sensor connected first to one of the USB interfaces and suffix 3 the sensor connected
second to one of the USB interfaces.
Return values:
<DEVice> 0 | 1 | OFF | ON
SENSe, READ and INITiate Subsystems