Remote Control Commands
R&S
®
SMC100A
282Operating Manual 1411.4060.32 ─ 09
*RST does not affect data lists. The numerical suffix at SOURce must not be used for
this command.
Parameters:
<Power> Power#1[, Power#2, ...]
Example:
CORR:CSET '/var/user/ucor1'
selects the table ucor1.
CORR:CSET:DATA:POW 1dB, 0.8dB, 0.75dB,...
enters the level values in the table ucor1.
Manual operation: See " Edit User Cor. Data - User Correction " on page 141
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
Queries the number of level values in the selected table.
The numerical suffix at SOURce must not be used for this command.
Return values:
<Points> integer
Range: 0 to 10000
*RST: 0
Example:
CORR:CSET '/var/user/ucor1'
selects the table ucor1.
CORR:CSET:DATA:POW:POIN?
queries the number of level values in the table ucor1.
Response: 440
the table ucor1 contains 440 level values.
Usage: Query only
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
The command fills the selected user correction list with the level values measured by
the power sensor for the given frequencies.
To select the used power sensor set the suffix in key word SENSe.
Example:
CORR:CSET:DATA:SENS:POW:SONC
fills the user correction list with level values acquired by the
power sensor connector to the SENSOR connector.
Usage: Event
Manual operation: See " Fill User Correction Data with Sensor" on page 146
[:SOURce]:CORRection:CSET:DELete <Filename>
Deletes the specified table.
SOURce Subsystem