Remote Control Commands
R&S
®
SMC100A
302Operating Manual 1411.4060.32 ─ 09
[:SOURce<hw>]:LFOutput:FREQuency:STOP <Stop>
Sets the stop frequency for the LF sweep.
Parameters:
<Stop> float
Range: full frequency range
Increment: see the data sheet: resolution of frequency setting
*RST: 100 KHz
Example:
LFO:FREQ:STOP 10 kHz
sets the stop frequency for the LF sweep to 10 kHz.
Manual operation: See " Stop Freq " on page 184
[:SOURce]:LFOutput[:STATe]
<State>
Activates/deactivates the LF output.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
LFO ON
activates the LF output. The settings under LFO:FREQ and
LFO:SWE become effective.
Manual operation: See " LF Output State" on page 180
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl <Dwell>
Sets the dwell time for each frequency step of the sweep.
Tip: It is recommended to switch off the "Display Update" for optimum sweep perform-
ance especially with short dwell times (SYSTem:DISPlay:UPDate OFF).
Parameters:
<Dwell> float
Range: see data sheet: Dwell time setting range
Increment: 100E-6
*RST: 15E-3
Example:
LFO:SWE:DWEL 20 ms
sets a dwell time of 20 ms.
Manual operation: See " Dwell Time - LF Sweep " on page 185
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute
Immediately starts an LF frequency sweep in LFO:SWE:MODE SINGle.
SOURce Subsystem