Remote Control Commands
R&S
®
SMC100A
316Operating Manual 1411.4060.32 ─ 09
Entering a level offset does not change the RF output level, but rather the query value
of :POWer.
For more information, see "RF level vs. RF output level" on page 129.
Only dB is permitted as the unit here. The linear units (V, W, etc.) are not permitted.
The keywords of this command are largely optional. Therefore, both the long and short
form of the command are shown in the example.
Parameters:
<Offset> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Example:
SOURce:POWer:LEVel:IMMediate:OFFSet -10
or
POW:OFFS 10
sets the RF level offset to 10 dB
Manual operation: See " Offset (Level) " on page 131
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL <Rcl>
Determines whether the RF level is retained or taken from a loaded instrument configu-
ration, when you recall instrument settings with the command *RCL.
Parameters:
<Rcl> INCLude | EXCLude
INCLude
Takes the level value of the loaded settings.
EXCLude
Retains the current level when an instrument configuration is
loaded.
*RST: INCLude
Example:
POW:RCL INCL
takes the level value from an instrument configuration loaded
with command *RCL.
Manual operation: See " Exclude Level " on page 114
[:SOURce<hw>]:POWer:LIMit[:AMPLitude] <Amplitude>
Limits the maximum RF output level in CW and SWEEP mode. It does not influence
the "Level" display or the response to the POW? query command.
The value is not affected by an instrument preset (PRESET key), *RST and the Save/
Recall function. This parameter is influenced only by the factory preset (SYST:FPR)
and its factory value is equal to the upper limit.
SOURce Subsystem