Remote Control Commands
R&S
®
SMC100A
318Operating Manual 1411.4060.32 ─ 09
Parameters:
<Manual> float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model
The values are given in the data sheet.
Increment: 0.01
*RST: -30
Example:
POW:SWE:MODE MAN
sets the Step sweep mode for RF output A.
POW:MAN -5 dBm
sets an RF level of -5 dBm for the next setting in the Step sweep
mode for RF output A.
POW:MODE SWE
sets the Level Sweep mode for RF output A.
POW:MAN -5.5 dBm
triggers the next sweep step with a level of -5.5 dBm.
Manual operation: See " Current Level - Level Sweep " on page 167
[:SOURce<hw>]:POWer:MODE <Mode>
Sets the instrument operating mode and therefore also the commands used to set the
output level.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW and FIXed are synonyms. To set the output level value, use
the command [:SOURce<hw>]:POWer[:LEVel][:
IMMediate][:AMPLitude].
SWEep
Operates in power sweep mode.
Set the range and current level with the commands [:
SOURce<hw>]:POWer:STARt, [:SOURce<hw>]:POWer:
STOP and [:SOURce<hw>]:POWer:MANual.
*RST:
CW
Example:
POW:MODE SWEep
selects the SWEep mode using the
POW:STAR; POW:STOP; POW:MAN settings.
Manual operation: See " State - Level Sweep " on page 164
[:SOURce<hw>]:POWer:POWer <Power>
Sets the RF level of the RF output connector.
SOURce Subsystem