Remote Control Commands
R&S
®
SMC100A
322Operating Manual 1411.4060.32 ─ 09
Parameters:
<Increment> float
Range: full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST: 1
Example:
POW:STEP 2
sets the step width for entering the RF level to 2 dB.
Manual operation: See " Variation Step " on page 133
[:SOURce<hw>]:POWer:STEP:MODE <Mode>
Activates (USER) or deactivates (DECimal) the user-defined step width used when
varying the level value with the level values UP/DOWN. The command is linked to set-
ting "Variation Active" for manual control, i.e. the command also activates/deactivates
the user-defined step width used when varying the level value with the rotary knob.
Parameters:
<Mode> DECimal | USER
*RST: DECimal
Example:
POW:STEP 2
sets the step width for the level setting to 2 dB.
POW:STEP:MODE USER
actives this step width for level variation with the rotary knob
(manual control) and with level values UP/DOWN (remote con-
trol).
Manual operation: See " Variation Active " on page 132
[:SOURce<hw>]:POWer:STOP <Stop>
Sets the stop level for the RF sweep.
Note: You can select any level within the setting range. The range is defined by the [:
SOURce<hw>]:POWer:STARt value and this stop value.
A defined offset ([:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet)
affects the level values according to the formula:
Minimum level + OFFSet ... Maximum level + OFFSet
Parameters:
<Stop> float
Determines the last level value of the sweep setting range.
Range: full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST: -10
SOURce Subsystem