Remote Control Commands
R&S
®
SMC100A
339Operating Manual 1411.4060.32 ─ 09
[:SOURce<hw>]:SWEep:POWer:DWELl <Dwell>
Sets the time taken for each level step of the sweep.
Tip: It is recommended to switch off the "Display Update" for optimum sweep perform-
ance especially with short dwell times (SYSTem:DISPlay:UPDate OFF).
Parameters:
<Dwell> float
Range: 1E-3 to 100
Increment: 100E-6
*RST: 15E-3
Example:
SWE:POW:DWEL 12 ms
sets a dwell time of 12 ms for a level sweep at the RF output.
Manual operation: See " Dwell Time - Level Sweep " on page 168
[:SOURce<hw>]:SWEep:POWer:EXECute
Triggers a sweep.
The command is only valid for sweep mode Single (SOURce:SWEep:POWer:MODE
SINGle). The command corresponds to the manual-control command "Execute Single
Sweep".
Example:
SOURce:SWEep:POWer:MODE SINGle
sets the single cycle mode of the level sweep.
SWE:POW:EXEC
triggers a level sweep at the RF output.
Usage: Event
Manual operation: See " Execute Single Sweep - Level Sweep " on page 166
[:SOURce<hw>]:SWEep:POWer:MODE <Mode>
Sets the cycle mode of the level sweep.
SOURce Subsystem