Remote Control Commands
R&S
®
SMC100A
343Operating Manual 1411.4060.32 ─ 09
[:SOURce<hw>]:SWEep:RESet[:ALL]
Resets all active sweeps to the starting point.
Example:
SWE:RES
resets all active sweeps to the starting point.
Usage: Event
Manual operation: See " Reset Sweep - Frequency Sweep " on page 160
7.14 STATus Subsystem
This system contains the commands for the status reporting system. See also Chap-
ter 6.5, "Status Reporting System", on page 223 for detailed information.
*RST on page 236 has no effect on the status registers.
Value ranges
●
Queries return the current value of the respective register, which permits a check of
the device status.
Return values: A decimal value in the range 0 to 32767 (=2
15
-1)
●
The configuration commands set the respective register thus determining which
status changes of the R&S SMC cause the status registers to be changed.
Setting values: A decimal value in the range 0 to 32767 (=2
15
-1)
:STATus:OPERation:CONDition...................................................................................... 343
:STATus:OPERation:ENABle.......................................................................................... 344
:STATus:OPERation[:EVENt].......................................................................................... 344
:STATus:OPERation:NTRansition....................................................................................344
:STATus:OPERation:PTRansition.................................................................................... 344
:STATus:PRESet........................................................................................................... 345
:STATus:QUEStionable:CONDition..................................................................................345
:STATus:QUEStionable:ENABle...................................................................................... 345
:STATus:QUEStionable[:EVENt]......................................................................................345
:STATus:QUEStionable:NTRansition................................................................................346
:STATus:QUEStionable:PTRansition................................................................................346
:STATus:QUEue[:NEXT]?............................................................................................... 346
:STATus:OPERation:CONDition <Condition>
Sets the content of the CONDition part of the STATus:OPERation register. This part
contains information on the action currently being performed in the instrument. The
content is not deleted after being read out because it indicates the current hardware
status.
Parameters:
<Condition> string
STATus Subsystem