EasyManua.ls Logo

Rohde & Schwarz SMC100A - Storing and Packing

Rohde & Schwarz SMC100A
413 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMC100A
356Operating Manual 1411.4060.32 ─ 09
Example:
SYSTem:COMMunicate:USB:RESource?
queries the VISA resource string for remote control via the USB
interface.
Response: "USB::72::000000::INSTR"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 96
:SYSTem:COMMunicate:SERial:RESource?
Queries the visa resource string for the serial remote control interface. This string is
used for remote control of the instrument.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:SERial:RESource?
queries the VISA resource string.
Response: "ASRL1::INSTR"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 96
:SYSTem:COMMunicate:SERial:BAUD <Baud>
Sets the baudrate for the serial remote control interface.
Parameters:
<Baud> 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200
*RST: 115200
Example:
SYSTem:COMMunicate:SERial:BAUD 115200
determines 115200 baudrate.
Manual operation: See "Baud Rate" on page 96
:SYSTem:COMMunicate:SERial:PARity <Parity>
Sets the parity for the serial remote control interface.
Parameters:
<Parity> NONE | ODD | EVEN
*RST: NONE
Example:
SYST:COMM:SER:PAR NONE
selects parity NONE.
Manual operation: See "Parity" on page 96