Digital Modulation SME
1038.6002.02 2.84 E-13
2.6.3.3.1 Recording a Data Sequence from an External Source
(External Loading)
Call DM menu À Mark one of the digital modulations using the menu cursor and press
key [SELECT].
Select external
À Mark parameter SOURCE using the menu cursor and press key
[SELECT].
À Mark selection EXT using the menu cursor and press key [SELECT].
Set the bit rate
À Mark submenu SELECT STANDARD... using the menu cursor and
press key [SELECT].
À Mark one of the standards using the menu cursor and press key
[SELECT].
or
À Mark parameter BITRATE using the menu cursor and press key
[SELECT].
À Enter the bit rate using rotary knob or numeric key and press key [1 x
ENTER].
Memory extension
À Mark submenu SELECT LIST... using the menu cursor and press key
[SELECT].
À Mark list XMEM using the menu cursor and press key [SELECT].
Define memory area
and memory
Note: In recording, the memory area is overwritten from the start area
and memory address to the stop address (START ADDRESS +
LENGTH –1)
Up to 7 bits below the start address and up to 15 bits above the
stop address (START ADDRESS + LENGTH –1) are additionally
overwritten with a random value. The additional bits must be
considered if a new recording is to be effected between
recordings already stored.
À Mark submenu CONFIG XMEM... using the menu cursor and press
key [SELECT].
À Mark parameter START ADDRESS using the menu cursor and press
key [SELECT].
À Enter the start address using rotary knob or numeric keys and press
key [1 x ENTER].
À Enter the start address using rotary knob or numeric keys and press
key [1 x ENTER].
À Mark parameter LENGTH using the menu cursor and press key
[SELECT].
À Enter the sequence length using rotary knob or numeric keys and
press key [1 x ENTER].
À Mark parameter MODE using the menu cursor and press key
[SELECT].
À Mark selection 8M*1 using the menu cursor and press key [SELECT].