Sweep SME
1038.6002.02 2.140 E-13
2.8.7 LEVEL Sweep
Menu SWEEP - LEVEL offers access to settings for LEVEL sweep.
Menu selection: SWEEP - LEVEL
FREQUENCY
LEVEL
MODULATION
DIGITAL MOD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
HELP
FREQ
LEVEL
LFGEN
FREQ
100. 000 000 0
MHz
START LEVEL
-30.0 dBm
STOP LEVEL -10.0 dBm
CURRENT LEVEL -30.0 dBm
STEP 1.0 dB
DWELL 15.0 ms
MODE OFF AUTO SINGLE STEP EXT-SINGLE EXT-STEP
EXECUTE SINGLE SWEEP
RESET SWEEP
MARKER 1 LEVEL +1.0 dB
MARKER 1 STATE OFF ON
MARKER 2 LEVEL +2.0 dB
MARKER 2 STATE OFF ON
MARKER 3 LEVEL +3.0 dB
MARKER 3 STATE OFF ON
START LEVEL
- 30.0
dBm
STOP LEVEL
- 10.0
dBm
LEV-SWP
Fig. 2-56 Menu SWEEP - LEVEL
START LEVEL
Input value of the starting level.
IEC bus short command :SOUR:POW:STAR -30dBm
STOP LEVEL
Input value of the stop levell. IEC bus short command
:SOUR:POW:STOP -10dBm
CURRENT LEVEL
Indication of the current level.
Operating mode STEP: Input value of the level.
STEP
Input value of the step width.
IEC bus short command :SOUR:SWE:POW:STEP 1dB
DWELL
Input value of the dwell time per step
IEC bus short command :SOUR:SWE:POW:DWEL 15ms
MODE
Selection of the sweep operating mode (cf. Section 2.8.3).
IEC bus short command :SOUR:POW:MODE SWE;
:SOUR:SWE:POW:MODE AUTO;
:TRIG:SOUR SING