Memory Sequence SME
1038.6002.02 2.150 E-13
External Trigger
An external signal at the rear input TRIGGER triggers the MEMORY SEQUENCE in the EXT-SINGLE
and EXT-STEP operating modes. The polarity of the active trigger edge can be set in the UTILITIES -
AUX I/O - EXT TRIG SLOPE menu.
Menu MEM SEQ with the two menu pages OPERATION and EDIT offers access to the
memory-sequence operating mode.
Menu selection: MEM SEQ
FREQUENCY
LEVEL
MODULATION
DIGITAL MOD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
HELP
MODE
RESET SEQUENCE
CURRENT INDEX
SELECT LIST...
DELETE LIST...
FUNCTION
OFF AUTO EXT-SINGLE EXT-STEP
STEP
CURRENT: MSEQ1
DELETE
FREQ
100. 000 000 0
MHz
LEVEL
-30.0
dBm
SINGLE
INSERT
FILL EDIT/VIEW
1
Fig. 2-61 Menu MEM SEQ -OPERATION-page (preset setting)
MODE
Selection of the operating mode; setting the operating mode
regards various command systems at the IEC bus (cf. above).
EXECUTE SINGLE SEQUENCE Ð
Starts the single run of a memory sequence. This menu option is
only visible if MODE SINGLE is selected.
IEC-bus short command :TRIG:MSEQ
RESET SEQUENCE Ð
Wrap to the beginning of the list.
IEC-bus short command :ABOR:MSEQ
CURRENT INDEX
Indication of the current list index. Setting value of the current list
index in the MODE STEP operating mode.
SELECT LIST...
Selection of a list or generation of a new list (cf. Section 2.2.4, List
Editor).
IEC-bus short command :SYST:MSEQ:SEL "MSEQ1"
DELETE LIST...
Deletion of a list (cf. Section 2.2.4, List Editor).
IEC-bus short command :SYST:MSEQ:DEL "MSEQ2"
FUNCTION
Selection of the editor functions to process a list (cf. Section 2.2.4,
List Editor).
IEC-bus short command :SYST:MSEQ 9,2,...;
:SYST:MSEQ:DWEL 50ms, 50ms,.