Utilities SME
1038.6002.02 2.156 E-13
EXT FREQUENCY
Input value of the external reference frequency (1 MHz to 16 MHz,
spacing 1 MHz).
IEC-bus short command :SOUR:ROSC:EXT:FREQ 10E6
ADJUSTMENT STATE
OFF Tuning value of the internal reference frequency as
calibrated (cf. menu UTILITIES-CALIB)
ON Tuning value according to setting value FREQUENCY
ADJUSTMENT. Option SM-B1, reference oscillator
OCXO, is switched off. Only the standard reference
oscillator is in operation.
IEC-bus short command :SOUR:ROSC:ADJ:STAT ON
FREQUENCY ADJUSTMENT
Input value in the range 0 to 4095 to set the internal reference
frequency. Pulling range ±4×10
-6
IEC-bus short command :SOUR:ROSC:ADJ:VAL 2048
2.11.6 Phase of the Output Signal (PHASE)
Menu UTILITIES-PHASE offers access to the phase setting of the RF output signal with respect to a
reference signal of the same frequency.
Menu selection: UTILITIES - PHASE
FREQUENCY
LEVEL
MODULATION
DIGITAL MOD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
HELP
SYSTEM
REF OSC
PHASE
PROTECT
CALIB
DIAG
TEST
MOD KEY
AUX I/O
BEEPER
INSTALL
DELTA PHASE
RESET DELTA PHASE DISPLAY
FREQ
100. 000 000 0
MHz
LEVEL
- 30.0
dBm
0 deg
Fig. 2-67 Menu UTILITIES - PHASE (preset setting)
DELTA PHASE
Setting value of the phase.
IEC-bus short command :SOUR:PHAS 0
RESET DELTA PHASE DISPLAY Ð
Sets the display of the DELTA PHASE to 0 without the phase of
the output signal being influenced.
IEC-bus short command :SOUR:PHAS:REF