CALibration SME
1038.6002.02 3.20 E-13
:CALibration:PULSe
The commands to calibrate the pulse generator are under this node (option SM-B4).
:CALibration:PULSe[:MEASure]?
The command triggers a calibration measurement. The command triggers an event and thus has
no *RST value.
Example: :CAL:PULS:MEAS? Response: 0
:CALibration:PULSe:DATA?
The command queries the correction data. It returns the correction data as two integers
separated by a comma. The first number indicates the fine adjustment, the second the coarse
adjustment.
Example: :CAL:PULS:DATA? Response: 26,2
:CALibration:ROSCillator
The commands to calibrate the reference oscillator are under this node.
:CALibration:ROSCillator[:DATA] 0 to 4095
The command enters the correction data. For an exact definition of the calibration value, cf.
Section 2.
Example: :CAL:ROSC:DATA 2048
:CALibration:VSUMmation
The commands to determine the support values for the frequency setting are under this node.
:CALibration:VSUMmation[:MEASure]?
The command triggers a calibration measurement. The command triggers an event and thus has
no default setting value.
Example: :CAL:VSUM:MEAS? Response: 0
:CALibration:VSUMmation:OFFS?
:CALibration:VSUMmation:DAC?
:CALibration:VSUMmation:KOS?
The commands query the calibration data (refer to service manual 1039.1856.24). They output all
calibration data in the format set in FORMat system.
Example: :CAL:VSUM:OFFS?