SOURce0|2:FREQuency SME
1038.6002.02 3.110 E-13
:SOURce0|2:FREQuency[:CW | :FIXed] 400 Hz | 1 kHz | 3 kHz | 15 kHz or 0.1 Hz to 500 kHz or
0.1 Hz to 1 MHz
The command sets the frequency for the CW mode.
If neither SM-B2 nor SM-B6 are fitted, the values 400 Hz, 1 kHz, 3 kHz and 15 kHz are
permissible for SOURce0. With option SM-B2, values from 0.1 Hz to 500 kHz are permissible,
with SM-B6 from 0.1 Hz to 1 MHz. RST-Wert is 1 kHz
Example: :SOUR2:FREQ:CW 1kHz
SOURce0|2:FREQuency:MANual 0.1 Hz to 500 kHz or 0.1 Hz to 1 MHz
The command sets the frequency if SOURce2:SWEep:MODe MANual and SOURce2:
FREQuency: MODe SWEep are set. In this case, only frequency values between the settings
SOURce2: FREQuency:STARt and to :STOP are allowed.
Example: :SOUR2:FREQ:MAN 1kHz *RST value is 1kHz
SOURce0|2:FREQuency:MODE CW|FIXed | SWEep
The command specifies the operating mode and hence by means of which commands the
FREQuency subsystem is checked. The following allocations are valid:
CW |FIXed CW and FIXed are synonyms. The output frequency is specified by means of
SOURce0|2:FREQuency:CW |FIXed.
SWEep The generator operates in the SWEep mode. The frequency is specified by means
of commands SOURce2:FREQuency:STARt; STOP; MANual. The SWEep
setting is only possible for SOURce2-
Example: :SOUR2:FREQ:MODE CW
*RST value is CW
SOURce0|2:FREQuency:STARt
0.1 Hz to 500 kHz or 0.1 Hz to 1 MHz
This command indicates the starting value of the frequency for the sweep.
Example:
:SOUR2:FREQ:STAR 100kHz *RST value is 1kHz
SOURce0|2:FREQuency:STOP
0.1 Hz to 500 kHz or 0.1 Hz to 1 MHz
This command indicates the end value of the frequency for the sweep.
Example:
:SOUR2:FREQ:STOP 200kHz *RST value is 100 kHz