SME SOURce2:SWEep
1038.6002.02 3.113 E-13
3.6.12.4 SOURce2:SWEep-Subsystem
This subsystem contains the commands to check the LF sweep of SOURce2. Sweeps are triggered on
principle.
Command Parameter Default
Unit
Remark
:SOURce2
:SWEep
:BTIMe
[:FREQuency]
:DWELl
:MODE
:POINts
:SPACing
:STEP
[:LINear]
:LOGarithmic
NORMal | LONG
1 ms to 1 s
AUTO | MANual | STEP
Number
LINear | LOGarithmic
0 to 500 kHz
0.01 PCT to 50 PCT
s
Hz
PCT
Option SM-B2
:SOURce2:SWEep:BTIMe NORMal | LONG
The command sets the blank time (Blank TIM
e) of the sweep. The setting is valid for all sweeps,
i.e., also for RF sweeps
NORMal Blank time as short as possible.
LONG Blank time long enough to permit an X/Y recorder to return to 0.
Example: :SOUR2:SWE:BTIM LONG *RST value is NORM
:SOURce2:SWEep
The commands to set the frequency sweeps are under this node. Keyword [:FREQuency] can be
omitted. Then the commands are SCPI-compatible unless stated otherwise (see examples).
:SOURce2:SWEep[:FREQuency]:DWELl 1 ms to 1 s
The command sets the time per frequency step (dwell).
Example: :SOUR2:SWE:DWEL 20ms *RST value is 15 ms
:SOURce2:SWEep[:FREQuency]:MODE AUTO | MANual | STEP
The command specifies the run of the sweep.
AUTOEach trigger triggers exactly one entire sweep cycle.
STEP Each trigger triggers only one sweep step (single-step mode). The frequency
increases by the value indicated under :SOURce2:SWEep
Example: :SOUR2:SWE:MODE AUTO *RST value is AUTO