3.6.15 TEST-System
This system contains the commands to execute the selftest routines (RAM?, ROM? and BATTery?) as
well as to directly manipulate the hardware modules (:TEST:DIRect). The selftests return a "0" if the test
has been executed successfully, otherwise a value unequal to "0". All commands of this system do not
have an *RST value.
Caution: The commands under node :TEST:DIRect directly act on the respective hardware module
circumventing any security mechanisms. They are provided for service purposes and
should not be used by the user. Improper use of the commands may damage the module.
Command Parameter Default
Unit
Remark
:TEST
:DIRect
:ATTC
:DCOD
:DSYN0MUX
:DSYN1MUX
:FMOD
:LFGENA
:LFGENB
:MGEN
:OPU1M
:OPU3M
:OPU6M
:PUM
:REFSS
:ROSC
:SUM
:RAM?
:ROM?
:BATTery[:RAM]?
:BATTery :XMEM?
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Subaddress, hex data string
Query only
Query only
Query only
Query only
:TEST:DIRect
This node contains the commands directly acting on the respective hardware module
circumventing any security mechanisms. The commands under this node have no short form.
:TEST:DIRect:SUM Subaddress, hex data string
The command directly acts on module SUM. A subaddress (0 or 1) must be indicated as a
parameter. The data are indicated as a <string> (i.e., an ASCII character string enclosed in
quotation marks) representing hexadecimal numbers. Thus characters 0 to 9 A to F may occur in
the character string.
Example: :TEST:DIR:SUM 0, "0010AF1F"
Query: :TEST:DIR:SUM? 0
:TEST:DIRect:ATTC Subaddress, hex data string
The command acts on module ATTC. (cf. :TEST:DIR:SUM)