SME Status-Reporting-System
1038.6002.02 3.142 E-13
3.8.5 Resetting Values of the Status Reporting Systems
Table 3-8 comprises the different commands and events causing the status reporting system to be
reset. None of the commands, except for *RST and SYSTem:PRESet influences the functional
instrument settings. In particular, DCL does not change the instrument settings.
Table 3-8 Resetting instrument functions
Event
Switching on
supply voltage
DCL,SDC
Power-On-Status-
Clear
(Device Clear,
Selected Device
Clear)
*RST or
SYSTem:PRESet
STATus:PRESet *CLS
Effect 0 1
Clear STB,ESR
yes

yes
Clear SRE,ESE
yes

Clear PPE
yes

Clear EVENt parts of the
registers
yes

yes
Clear ENABle parts of all
OPERation-and
QUESTionable registers,
Fill ENABle parts of all
other registers with "1".
yes

yes
Fill PTRansition parts with
„1"
Clear NTRansition parts
yes

yes
Clear error queue yes yes

yes
Clear output buffer yes yes yes 1) 1) 1)
Clear command
processing and input
buffer
yes yes yes

1) Every command being the first in a command line, i.e. immediately following a <PROGRAM MESSAGE TERMINATOR>
clears the output buffer.