SME Test Instruments and Utilities
1038.6002.02 5.1 E-13
5 Performance Test
5.1 Test Instruments and Utilities
Table 5-1 Test Instruments and Utilities
Item Type of instrument Specifications required Suitable R&S
instrument
R&S order no. Application
1 Frequency counter
(included item 2)
1Hz to 1.5 GHz (SME02)
1Hz to 3 GHz (SME03),
1Hz to 6 GHz (SME06),
Resolution 0.1 Hz
5.2.2
5.2.4
2 RF spectrum analyzer 100 Hz to 5 GHz (SME02/03),
100 Hz to 18 GHz (SME06)
synthesizer tuning,
dynamic range >80 dB
FSB
FSM
0848.0020.52
1020.7020.52
5.2.3
5.2.5
5.2.6
5.2.7
5.2.8
5.2.11
5.2.12
5.2.32
5.2.33
5.2.34
5.2.35
3 Storage oscilloscope 100-M samples/s,
averaging function
5.1.1.5
5.2.3
4 Controller Industrial standard PC/XT/AT
with IEC-625 interface
R&S Basic
PSA 15 1012.1003.03 5.2.3
5.2.35
5 Signal generator of high
spectral purity
0.1 MHz to 1.5 GHz (SME02)
0.1 MHz to 3 GHz (SME03),
0.1 MHz to 6 GHz (SME06),
SSB noise level <-126 dBc
with 1GHz/20 kHz
SME02,
SME03,
SME06
1038.6002.02/03/0
6
5.1.1.3
5.1.1.4
5.1.1.5
5.2.7
5.2.8
5.2.12
5.2.14
6 Phase noise test setup Mixer:
10 MHz to 1.5 GHz (SME02)
10 MHz to 3 GHz (SME03),
10 MHz to 6 GHz (SME06),
separating filter 2 MHz,
preamplifier with
40-dB changeover switch,
input noise <2nV (1Hz)
5.2.7
5.2.8
7 Oscilloscope Bandwidth > 100 MHz,
two channels with d.c. coupling
5.1.1.4
5.2.7
5.2.8
5.2.40
8 RF power meter 5 kHz to 1.5 GHz (SME02)
5 kHz to 3 GHz (SME03),
5 kHz to 6 GHz (SME06),
NRVS
NRVS-Z51
1020.1809.02
857.9004.02
5.2.11
5.2.13
9 Precision attenuation set Attenuation 0 to 120 dB,
resolution 5 dB
DPSP
RSG
8334.6010.02 5.2.11