EasyManua.ls Logo

Rohde & Schwarz SME02 - Output Reflection Coefficient

Rohde & Schwarz SME02
462 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SME Test Procedure
1038.6002.02 5.17 E-13
5.2.12 Output Reflection Coefficient
Test equipment -VSWR bridge (Section 5.1, item 11)
- VSWR directional coupler (Section 5.1, item 25) for SME06
- Second signal generator (Section 5.1, item 5)
- Spectrum analyzer (Section 5.1, item 2) .
Test setup
Note: For test frequencies >3 GHz use a VSWR directional coupler
instead of a VSWR bridge.
The test port of the bridge or the directional coupler has to be
screwed directly to the DUT.
×
Ø
S
M
E
RF
Signal generator
Source refl.
Object output
10-MHz reference
LO
Analyzer
Bridge (test frequency
3 GHz)
Directional coupler (test frequency > 3 GHz)
Measurement À Settings at the SME
-Test level
- Test frequency, unmodulated.
À Settings at the Spectrum analyzer
- Test frequency
-Test level
- Resolution and video bandwidth 10kHz,
- Span 0 Hz,
- linear scale
- Sweep time 30 ms.
À Settings at the second signal generator
- Detune by 100 Hz compared to the test frequency,
- first minimal level, unmodulated.
À At the spectrum analyzer, now bring the line indicated approx. into
the center of the screen by varying the reference level and read
and note down the level as a reference level.
À Screw off the bridge or the directional coupler from the SME and
increase the level at the second signal generator until the reference
level is measured again at the analyzer.

Table of Contents