Digital Standard PHS SMIQ
1125.5555.03 E-92.126
SELECT SLOT...
Selection of one of 8 possible slots. When selecting the slot, a window is
opened in which the data contents belonging to this slot can be defined.
4 slots are available for uplink and downlink. They are designated as UP<i>
and DN<i> in the display. For remote control, the slots are numbered from 1 to
8. The following assignment applies:
Slot DN1 DN2 DN3 DN4 UP1 UP2 UP3 UP4
Slot number 1 2 3 4 5 6 7 8
If the cursor is placed onto a slot in the diagram, it may be switched on and off
by pressing one of the unit keys (toggle function).
FREQ
100. 000 000 0
MHz
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
PHS
PHS
IS95
NADC
PDC
GSM
SLOT R SS PR UW CS_ID PS-ID IDLE CRC GUARD
1DN4262 32 42 28 34 16 16
BURST TYPE... SYNC
SLOT LEVEL OFF ATTEN FULL
SET DEFAULT
CS-ID SCRAMBLE STATE OFF ON
CS ID CODE 000 H
----------------BURST CONTENTS----------------
ROH
SS 2 H
PR 1999 9999 9999 9999 H
UW 50EF 2993 H
CI 9 H
CS ID 202 0002 0001 H
PS ID 000 0001 H
IDLE 0 0000 0000 H
CRC CI+CS_ID+PS_ID+IDLE
GUARD 0000 H
-----------------CHANGE DATA------------------
SELECT DATA LIST... CURRENT: R&STDM
COPY CURRENT DATA LIST TO...
DELETE DATA LIST...
EDIT DATA LIST...
LEVEL
-30.0
dBm
PEP
- 27.5
dBm
Fig. 2-75 Menu DIGITAL STD - PHS - SELECT SLOT, SMIQ equipped with Modulation Coder