SMIQ LIST Mode
1125.5555.03 2.403 E-9
MARKER
At the first step of the LIST mode, this output provides an approx. 200 µs
trigger signal immediately after blanking. At small DWELL times, this signal
can be used for an accurate synchronization to trigger other devices and
shows the first stable output frequency. The delay to the fed-in signal at the
TRIGGER input for EXT-SINGLE or EXT-STEP is 1.5 to 2 ms and has a jitter
of 0.5 ms.
TRIGGER
Input
BLANK
Output
MARKER
Output
Frequency
Fig. 2-249 Signal example LIST mode: MODE = EXT-STEP
The LIST menu offers access to settings for the LIST mode.
Menu selection: LIST
LEVEL
- - -.-
- - -.- - -
MODE
OFF AUTO SINGLE STEP EXT-SINGLE EXT-STEP
RESET LIST
DWELL 10.0 ms
CURRENT INDEX 1
LEARN
SELECT LIST... CURRENT: LIST2
DELETE LIST...
FUNCTION FILL INSERT DELETE EDIT/VIEW
FREQ
LIST
EXECUTE SINGLE LIST
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-250 Menu LIST - OPERATION page