Rohde & Schwarz SMIQ02B - Assigning Modulations to the [MOD ON;OFF] Key (MOD-KEY)

Rohde & Schwarz SMIQ02B
516 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Utilities SMIQ
1125.5555.03 E-92.428
2.29.15 Assigning Modulations to the [MOD ON/OFF] Key (MOD-KEY)
The modulations can be switched on/off in the individual modulation menus and parallely by means of
the [MOD ON/OFF] key.
For which modulations the [MOD ON/OFF] key is effective can be defined in the UTILITIES-MOD KEY
menu. The key can either be effective for all modulations or for a selected one.
Function of the [MOD ON/OFF] key if effective for a type of modulation:
Ø Every pressing a key alters the state (ON or OFF) of the selected modulation.
Function of the [MOD ON/OFF] key if effective for all types of modulation (ALL):
Ø If at least one modulation is switched on, pressing the [MOD ON/OFF] key switches the modulation/s
off. Which modulations were switched on is stored.
If no modulation is switched on, pressing the [MOD ON/OFF] key switches on the modulations which
were last switched off using the [MOD ON/OFF] key.
On switching on using the [MOD ON/OFF] key, the modulation sources are used as defined in the
modulation menus.
Access to the selection of the modulation to be switched using the [MOD ON/OFF] key is possible in the
UTILITIES-MOD KEY menu.
Menu selection: UTILITIES - MOD KEY
SYSTEM
REF OSC
PHASE
PROTECT
CALIB
DIAG
TEST
MOD KEY
AUX I/O
BEEPER
INSTALL
MODULATION... ALL
FREQ
100. 000 000 0
MHz
LEVEL
-30.0
dBm
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-272 Menu UTILITIES - MOD KEY (preset setting)
MODULATION
Open a window for selection of the modulation for which the [MOD ON/OFF]
key is to be effective. The following is to choose from:
ALL, AM, BB-AM, FM, PM, PULSE, DM, VM, FSIM, DISTO, NOISE.
Note: Preset switches off all modulations, sets the selection to ALL and
stores AM 30%, AM SOURCE INT as default setting.