SMIQ Utilities
1125.5555.03 2.429 E-9
2.29.16 Setting Auxiliary Inputs/Outputs (AUX-I/O)
Menu UTILITIES - AUX I/O offers access to settings for the TRIGGER input, BLANK output and
MARKER output. Sections Sweep, LIST Mode and Memory Sequence provide further information.
Menu selection: UTILITIES - AUX I/O
SYSTEM
REF OSC
PHASE
PROTECT
CALIB
DIAG
TEST
MOD KEY
AUX I/O
BEEPER
INSTALL
EXT TRIGGER SLOPE
POS NEG
SWEEP BLANK TIME NORM LONG
BLANK POLARITY NORM INV
MARKER POLARITY NORM INV
FREQ
100. 000 000 0
MHz
LEVEL
- 30.0
dBm
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-273 Menu UTILITIES - AUX I/O
EXT TRIGGER SLOPE
Selection of the active edge of the external trigger signal.
POS The instrument triggers on the positive edge of the
external signal.
NEG The instrument triggers on the negative edge of the
external signal.
IEC/IEEE-bus command :TRIG:SLOP POS
SWEEP BLANK TIME
Selection of the blank duration.
NORM BLANK duration is set to the shortest duration
possible.
LONG The BLANK duration is set for the PEN LIFT control of
an XY recorder (approx. 500ms).
IEC/IEEE-bus command :SOUR2:SWE:BTIM NORM
BLANK POLARITY
Selection of the polarity for the blank signal.
NORM With active BLANK, the output signal is HIGH.
INV Polarity is inverted.
IEC/IEEE-bus command :OUTP:BLAN NORM
MARKER POLARITY
Selection of the polarity for the marker signal.
NORM The output signal is HIGH when the sweep cycle
reaches the mark.
INV Polarity is inverted.
IEC/IEEE-bus command SOUR:MARK:POL NORM