SMIQ Error Messages
1125.5555.03 2.433 E-9
2.32 Error Messages
The SMIQ displays error and caution messages in a different manner, depending on how long, for a
short period of time or permanently, the cause exists.
Short-term message
The short-term message is displayed in the status line. Part of it
overwrites the status indications and disappears after approx. 2
seconds or in the case of a new entry.
The instrument shows, e.g., short-term messages if the attempt is
made to enter an overrange or if incompatible operating modes
deactivate one another.
Long-term message
The long-term message is displayed in the status line by means of
the message "WARNING" or "ERROR". Pressing the [ERROR] key
calls the ERROR page in which the messages are entered. Several
messages can be entered at the same time. The long-term message
remains existing until there is no cause any more. The ERROR page
is exited using the [RETURN] key.
The instrument displays, e.g., the long-term message "ERROR" if
there is a hardware error or "WARNING" if overrange settings have
been made.
Notes: An error message "ERROR" does not necessarily point to a defect instrument. There are
various operating states which can cause an ERROR message. E. g. if the instrument is set
to external reference but no external reference is connected.
Error 313 indicates the loss of calibration data and is also applicable in case of a cold start
(key [PRESET] is pressed during switch-on). The calibration values can be restored with
internal calibration routines. These routines are accessible via menu UTILITIES-CALIB (see
section on calibration).
The ERROR page offers access to long-term messages if the [ERROR] key is pressed.
WARNING -221 Settings conflict; modulation forces peak level into overrange
ERROR 211 Summing loop unlocked
WARNING 153 Input voltage out of range; EXT1 too low
FREQ
100. 000 000 0
MHz
AM
LEVEL
12.0
dBm
EXT1-LOW ERROR
Fig. 2-277 ERROR page
A list of the possible error messages is to be found in annex B.