SMIQ Fitting the Options
1125.5555.03 E-71.9
Fitting the option ½ Plug the module into the appropriate slot, lock it and fasten all
screws.
½ Open the air inlets at the housing frame by breaking out the safety
glass plate which belongs to the option.
½ Depending on the fact whether the option SMIQB20 (MCOD) has
been installed in the SMIQ, the following coaxial connections have
to be made:
Note: Please store the remaining cables. They will be required if
further options will be installed at a later stage or options will
be removed.
SMIQ with option MCOD SMIQ without option MCOD Remark about cable
Cable from FSIM1 to to
W361 X361 MCOD X325 - W361 was provided with FSIM1
W244 X361 - FRO I socket Withdraw W244 from X244 of module
IQMOD
W363 X363 MCOD X328 - W363 was provided with FSIM1
W245 X363 - FRO Q socket Withdraw W245 from X245 of module
IQMOD
W367 X367 IQMOD X244 IQMOD X244 W367 was provided with FSIM1
W388 X368 REAR I FADED REAR I FADED W388 was provided with FSIM1
W370 X370 IQMOD X245 IQMOD X245 W370 was provided with FSIM1
W391 X371 REAR Q FADED REAR Q FADED W391 was provided with FSIM1
X371 X370 X369 X368 X367 X366 X365
X360
X364 X363
X362 X361
FSIM1
Fig. 1-2 Module FSIM
½ The included adhesive label ’Option included’ is to be fixed at the
rear panel of the SMIQ.
Calibrating the
Fading Simulator and the ½Warm-up the instrument for 2 hours.
I/Q Modulator ½Call up menu UTILITIES - CALIB - VECTOR MOD.