Fitting the Options SMIQ
1125.5555.03 E-71.10
1.3.8 Option SMIQB15 - Second Fading Simulator (FSIM2)
The second Fading Simulator is fitted at the slot with label ’FSIM2’.
Before fitting the option SMIQB15 (FSIM2) please check the correct settings of the jumpers on both
fading modules FSIM1 (see Section 1.3.7) and FSIM2. The jumper settings depend on the series of the
FSIM module, they are also indicated on the screening cover.
After an instrument warm-up period of 2 hours, the IQ Modulator should be calibrated.
Checking the jumper Jumper Setting for the Fading Simulator FSIM2:
settings (cf. module number and labeling ’Jumper Setting’ on the screening
cover of the module)
FSIM2 FADING SIMULATOR
1085.XXXX
FADING SIMULATOR
1114.XXXX
Jumper connects X1.2 with X1.3 X1.1 with X1.2
X8.2 with X8.3 X8.2 with X8.3
X9.2 with X9.3 X9.1 with X9.2
X12.1 with X12.2 X12.1 with X12.2
X13.2 with X13.3 X13.2 with X13.3
X15.2 with X15.3 X15.2 with X15.3
X16.2 with X16.3 none
Fading modules of series 1114.XXXX also require the setting of a DIP switch. The setting is indicated
on the screening cover.
FADING SIMULATOR
1114.XXXX
Switch
12
FSIM1
OFF ON
FSIM2
ON OFF