R&S®SML / R&S®SMV03 Stereo Modulation (Option R&S SML-B5)
1090.3123.12 4.27 E-6
EON-DEL
Description: Enhanced Other Networks:
deletes the complete EON with PI=xxxx.
Command: EON-DEL=xxxx
Value range: xxxx = 0000 to FFFF (ASCII coded hexadecimal numbers))
Example: STEReo:DIRect “EON-DEL=1000“
Deletes the EON with PI=1000.
EON-PI
Description: Enhanced Other Networks:
creates a new EON or reads the list of the program identification (PI) codes of all
EONs created so far.
Command: EON-PI=xxxx
Query: EON-PI?
Response: xxxx,...
Value range: 0000 to FFFF (ASCII coded hexadecimal numbers)
Example: Command:
STEReo:DIRect “EON-PI=1000“
Creates a new EON with PI=1000.
Query:
STEReo:DIRect? EON-PI“
Response:
"1000"
Note: A four-digit value must always be set.
Leading zeros, if any, must also be specified.
A maximum of eight EONs can be created.
EON-PS
Description: Enhanced Other Networks:
sets the program service (PS) name for the EON with PI=yyyy.
Command: EON-PS=yyyy,xxxxxxxx
Query: EON-PS,yyyy?
Response: xxxxxxxx
Value range: xxxxxxxx = 8 ASCII characters
yyyy = 0000 to FFFF (ASCII coded hexadecimal numbers)
Example: Command:
STEReo:DIRect “EON-PS=1000,Test 123“
Sets the program service name for the EON with PI=1000 to "Test 123".
Query:
STEReo:DIRect? EON-PS,1000“
Reads the program service name of the EON with PI=1000.
Response:
"Test 123"
Note: An eight-digit value must always be set. Blank spaces, if any, must also be entered,
otherwise the value will not be accepted.