R&S SMR SAVE/RECALL – Storing/Calling of Instrument Settings
1134.9108.62 E-1 3.15
SAVE/RECALL – Storing/Calling of Instrument Settings
50 complete instrument settings can be stored in memory locations 1 to 50.
Operating Steps Explanations
SAVE
DATA INPUT
x1
ENTER
1
2
dB(m)
Store current instrument setting in
memory location 12.
RCL
DATA INPUT
x1
E
NTER
1
2
dB(m)
Call instrument setting of memory
location 12.
The digital display during a save or recall entry is faded in a window.
If an instrument setting is stored in which a sweep was switched on, the sweep is started using the
recall.
The parameter Exclude From Recall in the Frequency and Level-Level menus determines whether the
saved RF frequency and RF level are loaded when an instrument setting is loaded, or whether the
current settings are maintained.
Store IEC-bus command: "*SAV 12"
Call IEC-bus command: "*RCL 12"
Note: The contents of lists, as they are used for the List mode or for user correction (Ucor), is not
saved in the SAVE memory. It is stored under the respective list name and can be called. If
instrument settings are called which go back to list data such as level setting using Ucor,
the current list contents is used. If this has been altered, it is not identical to the list
contents at the point of storing any more.