Instrument Function
R&S
®
SMU200A
227Operating Manual 1007.9845.32 ─ 15
"Insert Range"
Insert new rows before the marked row. The number of rows to be inser-
ted can be defined in an entry window.
"Fill...."
Opens a sub menu for defining a set of list values to be automatically
entered in the ucor list (see chapter 5.5.9.2, "Filling the Correction List
automatically", on page 229).
"Fill With Sen-
sor"
Calls the menu to activate the filling of the user correction list with level
values acquired by the selected power sensor (see chapter 5.5.9.3, "Fill-
ing the Correction List with Power Sensor Measurement Data",
on page 230).
"Delete Row"
Deletes the marked row.
"Delete
Range..."
Allows to delete any number of rows starting with the marked row. The
number of rows to be deleted can be defined in an entry window.
"Save as"
Open the file menu to save the list under a new name.
Note: Save list only after filling both columns (frequency and level), oth-
erwise the entries are lost.
Each list is saved to the R&S Signal Generator hard disk as a separate
file with the file prefix *.uco. The file name and the directory to which
the file is saved are user-selectable.
"Save"
The list is saved under its current name.
SCPI command:
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 706
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency on page 699
[:SOURce<hw>]:CORRection:CSET:DATA:POWer on page 700
Import/Export
User correction list can be imported from externally created files or exported into text or
CSV-files. The import/export settings are available after clicking the "Import/Export" but-
ton.
Import/Export - User Correction
Expands the menu with the area for import and export of user correction files.
Externally edited Excel tables with any number of frequency/level value pairs can be
imported as text or CSV-files and used for user correction.
Multi level lists which are filled automatically using the "Fill with sensor" function consist
of nine level columns, one for each attenuator stage (see chapter 5.5.9.3, "Filling the
Correction List with Power Sensor Measurement Data", on page 230).
On the other hand, internally created user correction list can be exported as text or CSV-
files.
RF Signal and Analog Modulations - RF Block