Instrument Function
R&S
®
SMU200A
258Operating Manual 1007.9845.32 ─ 15
"Extern Start/
Stop"
Sets an automaticly repeated sweep cycle that is started, stopped and
restartet by subsequent external trigger events.
The first external trigger signal starts the sweep (Start).
The next external trigger signal stops the sweep at the current frequency
(Stop).
The third external trigger signal starts the sweep at the start frequency
(Start).
If a different sweep mode was activated prior to the "Extern Start/Stop"
mode, the current sweep is stopped and the "Extern Start/Stop" sweep
starts at the start frequency when triggered.
The "Reset" button resets the sweep to the start level.
The external trigger signal is input at the rear of the instrument (BNC
connector INST TRIG).
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR EAUT
SOUR:POW:MODE SWE (External trigger)
SCPI command:
[:SOURce<hw>]:SWEep:POWer:MODE on page 781
:TRIGger<hw>:PSWeep:SOURce on page 801.
[:SOURce<hw>]:POWer:MODE on page 768
Reset Sweep - Level Sweep
Resets the sweep. The start level is set and the next sweep starts from there.
SCPI command:
[:SOURce<hw>]:SWEep:RESet[:ALL] on page 784
Execute Single Sweep - Level Sweep
Triggers the sweep manually. A manual sweep can only be triggered if "Mode Single" is
selected.
SCPI command:
[:SOURce<hw>]:SWEep:POWer:EXECute on page 781
Start Level - Level Sweep
Sets the start level.
SCPI command:
[:SOURce<hw>]:POWer:STARt on page 770
Stop Level - Level Sweep
Sets the stop level.
SCPI command:
[:SOURce<hw>]:POWer:STOP on page 771
Current Level - Level Sweep
Displays the current level.
RF Signal and Analog Modulations - RF Block