"Single"
Single cycle from the beginning to the end of the list. If "Single" is
selected, the cycle is not started immediately. The "Execute Single" but-
ton appears under the "Mode" line. The cycle is started with this button.
The duration of a list step is determined by the set dwell time.
Button "Reset" restarts the list at the starting point.
"Step"
Manual, step-by-step processing of the list. Activating "Step" stops the
current list and the cursor moves to the value displayed for "Current
Index". It is now possible to scroll up and down in the list in discrete steps
by varying the index. The duration of a list step is determined by the time
between two index entries.
Button "Reset" restarts the list at the starting point.
"Extern Single"
Single cycle from the beginning to the end of the list as with "Single", but
started by an external trigger.
The external trigger signal is input at the BNC connector INST TRIG.
The BNC connector INST TRIG is provided for Path A, and the pin INST
TRIG B on the AUX I/O interface for Path B.
Button "Reset" restarts the list at the starting point.
"Extern Step"
Step-by-step cycle using the external trigger signal. Each trigger event
starts a single step. The duration of a list step is determined by the time
between two trigger events.
The external trigger signal is input at the BNC connector INST TRIG.
The BNC connector INST TRIG is provided for Path A, and the pin INST
TRIG B on the AUX I/O interface for Path B.
Button "Reset" restarts the list at the starting point.
SCPI command:
[:SOURce<hw>]:LIST:MODE on page 756
[:SOURce<hw>]:LIST:TRIGger:SOURce on page 759
Execute Single - List Mode
Triggers the list manually. This button is available only if "Mode Single" is selected.
SCPI command:
[:SOURce<hw>]:LIST:TRIGger:EXECute on page 758
Reset - List Mode
Resets the list to the starting point.
SCPI command:
[:SOURce<hw>]:LIST:RESet on page 758
Dwell Time - List Mode
Enters the dwell time. The dwell time determines the duration of a list step in list operating
modes "Auto", "Single" and "Extern Single". In these modes a complete list is processed
either once or continuously.