Instrument Function
R&S
®
SMU200A
352Operating Manual 1007.9845.32 ─ 15
"External Paral-
lel (AUX I/O)"
External parallel data is supplied externally via the PARDATA pins on
the AUX I/O interface.
SCPI command:
[:SOURce<hw>]:BB:DM:SOURce on page 622
List Management - Digital Modulation
Calls the menu for managing data and control lists (see chapter 5.8.3.6.8, "List Manage-
ment", on page 359).
SCPI command:
n.a.
Standard, Symbol Rate, Coding
The middle part of the menu is used for selecting the standard, the symbol rate and the
coding.
Set acc. to Standard - Digital Modulation
Selects a standard.
After selection, modulation parameters "Modulation Type", "Symbol Rate", "Filter" and
"Coding" are automatically set in accordance with the standard.
If one of these parameters is subsequently altered, the display changes to User. The User
setting can be saved to a file so that it can be recalled at some later time ("Save/Recall
User.." button).
The following table shows the standards that are available, together with the associated
settings of the modulation parameters.
Table 5-11: Standards - Custom Digital Modulation
Standard Modulation Symbol Rate Filter Coding Parameter for
IEC command
Bluetooth 2FSK, Devia-
tion 160.0 kHz
1.0 Msym/s Gauss, B*T =
0,5
OFF BLUetooth
DECT 2FSK, Devia-
tion 288.0 kHz
1.152 Msym/s Gauss, B*T =
0,5
OFF DECT
ETC (ARIB STD
T55)
ASK, ASK
Depth 100%
1.024 Msym/s Split Phase, B*T
= 2.0
OFF ETC
GSM MSK 270.833333
ksym/s
Gauss, B*T =
0.3
GSM GSM
GSM EDGE 8PSK EDGE
(3pi/8 8PSK)
270.833333
ksym/s
Gauss linear OFF GSMEdge
Baseband Signal - Baseband Block