Remote Control Commands
R&S
®
SMU200A
494Operating Manual 1007.9845.32 ─ 15
Query onlyUsage:
*WAI
WAIt to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
EventUsage:
7.2
Preset Commands
The preset commands are not bundled in one subsystem. Therefore, they are listed sep-
arately in this section. In addition, a specific preset command is provided for each digital
standard and for the fader. These specific commands are described in the associated
subsystems.
Four presetting actions are available:
●
Activating the default state of all internal instrument functions (*RST on page 492).
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. GPIB address or reference oscillator source settings.
●
Activating the preset state of the parameters related to the selected signal path
(:SOURce<hw>:PRESet on page 494)
●
Activating the preset state of all parameters that are not related to the signal path
(:DEVice:PRESet on page 494)
●
Activating the original state of delivery (factory reset, :SYSTem:FPReset
on page 495). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
:DEVice:PRESet
The command presets all parameters which are not related to the signal path. This
includes presetting the LF generator and bit/block error measurement.
DEV:PRES
presets all instruments settings that are not related to the signal
path
Example:
EventUsage:
:SOURce<hw>:PRESet
The command presets all parameters which are related to the selected signal path.
Fading (if available) and transient recorder are only preset by command *RST.
Preset Commands