Preparing for Use
R&S
®
SMU200A
37Operating Manual 1007.9845.32 ─ 15
The PRESET key calls up a defined instrument setup. All parameters and switching
states are preset (also those of inactive operating modes). The default instrument
settings provide a reproducible initial basis for all other settings. However, functions
that concern the integration of the instrument into a measurement setup are not
changed, e.g. IEC/IEEE bus address or reference oscillator source settings.
In remote control it is possible to preset the settings of the two instrument's paths
separately (see section "Preset Commands").
Preset the instrument to its factory settings
The instrument can also be forced to load its default factory settings. To access the
corresponding dialog box, press the SETUP key and select the "Factory Preset".
For more information and an overview of the settings affected by the factory preset
function, see chapter 5.2.3.18, "Factory Preset", on page 148.
Overview of the Most Important Preset States
The following list gives an overview of the presets for the most important generator set-
tings. The other presets can be found in the preset tables of the individual menus and
the information accompanying the IEC/IEEE-bus commands.
"RF frequency" = 1 GHz
"RF level" RF output switched off
"Offsets" = 0
"Modulations State" = Off
Uninterrupted level setting are switch off
"Level Attenuator Mode" = AUTO
Internal level control "Level ALC" = AUTO
User correction "Level Ucor" = OFF
"LF output State" = Off
"Sweep State" = Off
"List mode State" = Off
Settings that are not affected by the PRESET key
Reference frequency settings ("Ref Oscillator" menu)
Power on settings ("Level/EMF" menu)
Network settings ("Setup" menu)
GPIB address ("Setup" menu)
Password and settings protected by passwords ("Setup" menu)
Start/Stop Gui Update ("Setup" menu)
Display and keyboard settings ("Setup" menu)
User-defined instrument states can be stored and called up in the "File" dialog.