Remote Control Commands
R&S
®
SMU200A
572Operating Manual 1007.9845.32 ─ 15
Parameters:
float
-50.00 to 40.00
0.01
0
dB
<Cnratio>
Range:
Increment:
*RST:
Default unit:
AWGN:CNR 10
sets a carrier/noise ratio of 10 dB.
Example:
[:SOURce<hw>]:AWGN:DISP:MODE <Mode>
Selects the display mode to Bseband or RF.
Parameters:
RF|BB
RF
<Mode>
*RST:
AWGN:DISP:MODE RF
sets the display mode to RF
Example:
[:SOURce<hw>]:AWGN:DISP:ORESults <Oresults>
This command selects the display of output results for the analog (DACIF) or the digital
(BBOUT) signal path.
Note: This command is only available with the option R&S SMU-B18, Baseband Digital
I/Q Out. If the instrument is not equipped with this option the parameters of the analog
signal path are displayed.
This command is available for modes Additive Noise and Noise Only (SOUR:AWGN:MODE
ADD|ONLY) and for Display Mode set to Baseband (AWGN:DISP:MODE BB).
Parameters:
ANALog|DIGital
ANALog
The parameters of the analog signal path are displayed.
DIGital
The parameters of the digital signal path are displayed.
ANALog
<Oresults>
*RST:
AWGN:MODE ONLY
activates the generation of a pure noise for path A.
AWGN:DISP:MODE BB
sets the display mode to Baseband.
AWGN:DISP:MODE DIG
displays the output results of the digital signal of path A.
Example:
SOURce Subsystem