Remote Control Commands
R&S
®
SMU200A
588Operating Manual 1007.9845.32 ─ 15
BB:ARB:TRIG:SOUR EXT
sets external triggering via the TRIGGER 1 connector.
BB:ARB:TRIG:MODE ARET
selects the Armed_Retrigger mode.
BB:ARB:TRIG:RMOD?
queries the current status of waveform output.
Response: RUN
the waveform is output, an external trigger was executed.
Example:
Query onlyUsage:
[:SOURce<hw>]:BB:ARB:TRIGger:SLENgth <Slength>
The command defines the length of the signal sequence to be output in the Single trigger
mode. The unit is defined with command SOUR:BB:ARB:TRIG:SLUNit. It is possible
to output deliberately just part of the waveform, an exact sequence of the waveform, or
a defined number of repetitions of the waveform.
Parameters:
float
1 to 2^32-1
1 Waveform length
sample
<Slength>
Range:
*RST:
Default unit:
BB:ARB:SEQ SING
sets trigger mode Single.
BB:ARB:TRIG:SLUN SAMP
sets unit Samples for the entry of sequence length.
BB:ARB:TRIG:SLEN 200
sets a sequence length of 200 samples. The first 200 samples of
the current waveform will be output after the next trigger event.
Example:
[:SOURce<hw>]:BB:ARB:TRIGger:SLUNit <Slunit>
The command defines the unit for the entry of the length of the signal sequence
(SOUR:BB:ARB:TRIG:SLEN) to be output in the "Single" trigger mode
(SOUR:BB:ARB:SEQ SING).
Parameters:
SAMPle|SEQuence
SEQuence
<Slunit>
*RST:
BB:ARB:SEQ SING
sets trigger mode Single.
BB:ARB:TRIG:SLUN SEQ
sets unit Sequence length for the entry of sequence length.
BB:ARB:TRIG:SLEN 2
sets a sequence length of 2 waveforms. The current waveforms
will be output twice after the next trigger event.
Example:
SOURce Subsystem