Remote Control Commands
R&S
®
SMU200A
592Operating Manual 1007.9845.32 ─ 15
Marker Settings
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:MODE....................................................592
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:OFFTime................................................592
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:ONTime..................................................593
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:PATTern.................................................593
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:PULSe:DIVider........................................593
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:PULSe:FREQuency.................................594
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:MODE <Mode>
The command defines the signal for the selected marker output.
Parameters:
UNCHanged|RESTart|PULSe|PATTern|RATio|TRIGger
UNCHanged
A marker signal as defined in the waveform file (tag 'marker mode
x') is generated.
RESTart
A marker signal is generated at every waveform start.
PULSe
A pulsed marker signal is generated. The pulse frequency (=
symbol rate/divider) is defined with the
SOUR:BB:ARB:TRIG:OUTP:PULS:DIV command and can be
queried with the SOUR:BB:ARB:TRIG:OUTP:PULS:FREQ?
command.
PATTern
A marker signal is generated with the aid of a user-definable bit
pattern. The bit pattern is entered with the aid of
command :BB:ARB:TRIGger:OUTPut:PATTern. The bit
pattern is a maximum of 32 bits long.
RATio
A regular marker signal corresponding to the Time Off / Time On
specifications in the
commands :ARB:TRIGger:OUTPut:OFFTime
and :ARB:TRIGger:OUTPut:ONTime is generated.
RESTart
<Mode>
*RST:
BB:ARB:TRIG:OUTP2:MODE PULS
selects the pulsed marker signal on output MARKER 2.
BB:ARB:TRIG:OUTP2:STAT ON
activates the pulsed marker signal on output MARKER 2.
Example:
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:OFFTime <Offtime>
SOURce Subsystem