[:SOURce<hw>]:BB:ARB:TSIGnal:SINE:PHASe <Phase>
The command sets the phase offset of the sine wave on the Q channel relative to the
sine wave on the I channel.
Parameters:
float
-180.00 DEG to +180.00 DEG
0.01 DEG
90 DEG
<Phase>
Range:
Increment:
*RST:
BB:ARB:TSIG:SINE:PHAS 90
sets a phase offset of 90 degrees.
Example:
[:SOURce<hw>]:BB:ARB:TSIGnal:SINE:SAMPles <Samples>
The command sets the sample rate for the sine signal in samples per period.
The resulting clock rate must not exceed the maximum ARB clock rate of 100 MHz.
The maximum value is automatically restricted by reference to the set frequency and has
to fulfill the rule Frequency * Samples <= ARB clock rate.
Parameters:
float
1
100
100 samples per period
<Samples>
Increment:
*RST:
Default unit:
BB:ARB:TSIG:SINE:SAMP 100
sets a sample rate of 100 samples per period.
Example:
SOURce:BB:ARB:MCAR Subsystem
The MCARrrier subsystem contains the commands for setting the Multi Carrier Wave-
form Generator.
Section chapter 7.3, "R&S Signal Generator Waveform and List Format", on page 496
describes the ARB waveform format and how it is used to transmit waveforms via the IEC
bus.
CARRier<ch>
The numerical suffix under CARRier distinguish between the carriers. The value range
is 0 .. 31.
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier:COUNt........................................................598
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier:SPACing.....................................................599
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:DELay.................................................599
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:FILE....................................................599
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:PHASe.................................................600
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:POWer.................................................600