EasyManua.ls Logo

Rohde & Schwarz SMU200A - Page 628

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
608Operating Manual 1007.9845.32 ─ 15
MMEM:CDIR 'D:/user/waveform'
sets the default directory to D:/user/waveform.
BB:ARB:MCAR:OFIL 'mcar1_2'
defines the file name mcar1_2.wv for the multi carrier waveform
file
Example:
[:SOURce<hw>]:BB:ARB:MCARrier:PRESet
The command sets all ARB multi carrier parameters to their default values.
BB:ARB:MCAR:PRES
resets the ARB multi carrier parameters to default values.
Example:
EventUsage:
[:SOURce<hw>]:BB:ARB:MCARrier:SAMPles?
The command queries the resulting file size. The file size is returned in samples.
Return values:
float<Samples>
BB:ARB:MCAR:SAMP?
queries the file size of the currently calculated multi carrier wave-
form.
Example:
Query onlyUsage:
[:SOURce<hw>]:BB:ARB:MCARrier:SETTing:CATalog?
This command queries the available settings files in the specified default directory. The
settings files are used to set the ARB multi carrier submenu. Only files with the file exten-
sion *.arb_multcarr will be listed.
Return values:
string<Catalog>
MMEM:CDIR 'D:/user/waveform'
sets the default directory to D:/user/waveform.
BB:ARB:MCAR:SETT:CAT?
reads out all the settings files in the default directory.
Response: mcar1, mcar2
the directory contains the configuration files mcar1.arb_mult
carr and mcar2.arb_multcarr.
Example:
Query onlyUsage:
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMU200A

Related product manuals