Remote Control Commands
R&S
®
SMU200A
631Operating Manual 1007.9845.32 ─ 15
Parameters:
float
1 to (2^32-1) Symbols
1000 symbols
symbol
<Slength>
Range:
*RST:
Default unit:
BB:DM:SEQ SING
sets trigger mode Single.
BB:DM:TRIG:SLEN 200
sets a sequence length of 200 symbols. 200 symbols will be output
after the next trigger event.
Example:
[:SOURce<hw>]:BB:DM:TRIGger:SOURce <Source>
The command selects the trigger source.
Parameters:
INTernal|EXTernal | BEXTernal|OBASeband
INTernal
Triggering is executed by means of the Trigger command
SOURce:BB:DM:TRIGger:EXECute,
TRIGger:BB:DM:IMMediate or *TRG in the case of remote
control and by means of "Execute Trigger" in the case of manual
operation.
EXTernal
Triggering is executed by means of the signal on the TRIGGER 1
connector.
BEXTernal
Triggering is executed by means of the signal on the TRIGGER 2
connector.
OBASeband
Triggering is executed by means of the trigger signal from the
second path.
<Source>
BB:DM:TRIG:SOUR EXT
sets external triggering via the TRIGGER 1 connector.
Example:
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:DELay <Delay>
The command specifies the trigger delay (expressed as a number of symbols) for external
triggering.
SOURce Subsystem