Remote Control Commands
R&S
®
SMU200A
632Operating Manual 1007.9845.32 ─ 15
Parameters:
float
0 to 2^16-1 Symbols
1 symbol for :BB:DM:CLOC:SOUR EXT, 0.01 sym-
bols for :BB:DM:CLOC:SOUR INT
0 symbols
symbol
<Delay>
Range:
Increment:
*RST:
Default unit:
BB:DM:TRIG:SOUR EXT
selects an external trigger via the TRIGGER 1 connector.
BB:DM:TRIG:DEL 50
sets a delay of 50 symbols for the trigger.
Example:
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:INHibit <Inhibit>
The command specifies the number of symbols by which a restart is to be inhibited fol-
lowing a trigger event. This command applies only in the case of external triggering.
Parameters:
float
0 to 2^26-1 Symbols
1 symbol
0 symbols
symbol
<Inhibit>
Range:
Increment:
*RST:
Default unit:
BB:DM:TRIG:SOUR EXT
selects an external trigger via the TRIGGER 1 connector.
BB:DM:TRIG:INH 200
sets a restart inhibit for 200 symbols following a trigger event.
Example:
[:SOURce<hw>]:BB:DM[:TRIGger]:SEQuence <Sequence>
The command selects the trigger mode.
SOURce Subsystem