EasyManua.ls Logo

Rohde & Schwarz SMU200A - Page 663

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
643Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:BB:DM:PRAMp:BBONly[:STATe] <State>
This command selects power ramping in the baseband only or mixed power ramping in
the baseband and the RF section.
The ON setting is mandatory if, with power ramping active, only the baseband signal is
output (I/Q outputs)
In case of two-path instruments, the ON setting is also mandatory if a baseband signal
is applied to two RF paths (RF A and RF B).
Only then can a signal with a defined, predictable level be output.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
BB:DM:PRAM:BBON ON
selects power ramping in the baseband only.
Example:
[:SOURce<hw>]:BB:DM:PRAMp:FDELay <Fdelay>
The command sets the delay in the falling edge. A positive value gives rise to a delay
and a negative value causes an advance. The setting is expressed in symbols.
Parameters:
float
-4.0 to +4.0
0.1
0
symbol
<Fdelay>
Range:
Increment:
*RST:
Default unit:
BB:DM:PRAM:FDEL 1
The falling edge starts 1 symbol later.
Example:
[:SOURce<hw>]:BB:DM:PRAMp:RDELay <Rdelay>
The command sets the delay in the rising edge. A positive value gives rise to a delay and
a negative value causes an advance. The setting is expressed in symbols.
Parameters:
float
-4.0 to +4.0
0.1
0
symbol
<Rdelay>
Range:
Increment:
*RST:
Default unit:
BB:DM:PRAM:RDEL 1
The rising edge starts 1 symbol later.
Example:
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMU200A

Related product manuals