[:SOURce<hw>]:BB:MCCW:CARRier:PHASe <CarrierIndex>,<Phase>
[:SOURce<hw>]:BB:MCCW:CARRier:PHASe? <CarrierIndex>
The command sets the start phase of the selected carrier.
The phase settings are only valid if optimization of the crest factor is disabled
(:SOURce:BB:MCCW:CFACtor:MODE OFF).
Setting parameters:
float
Sets the start phase of the selected carrier.
0 DEG to 360 DEG
0.01 DEG
0 DEG
Parameters for setting and query:
float
Determines the carrier number.
0 to Max Number of Carriers
<Phase>
Range:
Increment:
*RST:
<CarrierIndex>
Range:
BB:MCCW:CARR:PHAS 15, 90 DEG
sets a start phase of 90 DEG for carrier 15.
Example:
[:SOURce<hw>]:BB:MCCW:CARRier:POWer <CarrierIndex>,<Power>
[:SOURce<hw>]:BB:MCCW:CARRier:POWer? <CarrierIndex>
The command sets the power of the selected carrier.
Setting parameters:
float
Sets the power of the selected carrier.
-80 dB to 0 dB
0.01 dB
0 dB
Parameters for setting and query:
float
Determines the carrier number.
0 to Max Number of Carriers
<Power>
Range:
Increment:
*RST:
<CarrierIndex>
Range:
BB:MCCW:CARR:POW 15, -50 dB
sets the power of carrier 15 to -50 dB.